User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 3400A:RMS Voltmeter DATE: 31-Aug-94 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 25 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- F W 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP Allow the UUT a 30 minute warm-up before testing. 1.005 DISP Switch the Test Instrument Power Off. 1.006 DISP Zero the meter movement using the center screw. 1.007 DISP Turn the Test Instrument on. 1.008 CALL HP 3400A Sub Procedure 1 1.009 CALL HP 3400A Sub Procedure 2 1.012 END